AIIC AI Intelligence Centre

SOURCE-LINKED INTELLIGENCE

Repeatability Characterisation and Error Budget of a Consumer Structured-Light Scanner for 3-D Wound Geometry:A Rigid-Phantom Study

arXiv · AI, language, vision and robotics · article · Aug 31, 2026 · UTC

We characterise the measurement error of a free- hand consumer structured-light scanner used to derive three- dimensional geometric wound descriptors. Rigid wound-care phantoms cannot change, so every difference between repeat scans of one site is measurement error; all repeats come from a single scanner unit, nine sites and 23 scans, so this charac- terises one instrument. The 95 percent repeatability limit for reconstructed surface area is a factor of 4.8, with a confidence interval from 3.0 to 7.0, so rescanning an unchanged site can shift the reading from a 79 percent decrease to a 377 per

Read original source ↗ Open in workspace

recordType
paper
region
Global

Evidence & attribution

First collected: 2026-09-21T07:22:03.933Z. This is not the publication date.