A Controlled Comparison of Manual and Teleoperated Intraocular Instrument Motion for an Input Device
Input devices for robotic microsurgery are frequently described as preserving the surgeon's trained technique, but the claim is rarely measured. We compared manual and teleoperated intraocular instrument motion with the trocar constraint, the instrument, the eye model and the tracking source common to both conditions, so that the control interface was the only factor varied. Prior comparisons cannot hold the instrument fixed, because a robotic instrument is not the tool used manually. Sixteen participants performed a navigation task on a commercial ophthalmic simulator by hand and through a th